September 2025

Journal

Thermocouples in Resistive and Induction Furnaces Operated in Strong Magnetic Fields

By:
Tener, Zachary P; Flynn, Steven; Donald, Alexander; Hamlin, James; Meisel, Mark; Manuel, Michele; Kesler, Michael S
Journal Name:
IEEE Transactions on Instrumentation and Measurement
Page Number:
1-5
Volume:
74
Publication Date:
September 18, 2025
View DOI Listing:
https://doi.org/10.1109/TIM.2025.3602597

Abstract

The thermomagnetic processing methods require an understanding of the in situ temperatures experienced by the workpieces. When commonly used thermocouples (TCs) are employed in induction furnaces, RF fields can contribute additional temperature uncertainties to the ones arising from the use of strong magnetic fields. Focusing on temperatures between 300 ∘ C and 1000 ∘ C produced by induction and resistive furnaces, the readings generated by Type-K TCs were contrasted to the ones produced by Type-S and Type-N sensors for magnetic fields up to 9 T. Overall, when comparing Type-K response to temperatures above 700 ∘ C in both zero field and high field ( ≤9 T), the differences amounted to less than 1% and when continuously measured has a linear relationship to the strength of the applied field. The relative invariance of Type-N and Type-S TCs was confirmed. These findings suggest that the use of TCs in high magnetic fields remains a viable option for applications, although the precision depends on the type used.


Related Researchers