Assessment of Low-Level Pu Isotope Ratio Measurements Using Multicollector Inductively Coupled Plasma Mass Spectrometry (MC-ICP-MS/MS) Equipped with a Pre-Mass Filter

By:
Zirakparvar, Nasser A; Manard, Benjamin T; Darnell, Matthew C; Adkisson, Melissa L; Metzger, Shalina C; Ticknor, Brian W; Hexel, Cole R; Dunlap, Daniel R
Journal Name:
Applied Spectroscopy Practica
Page Number:
1-25
Volume:
3
Issue Number:
3
Publication Date:
September 15, 2025
View DOI Listing:
https://doi.org/10.1177/27551857251376027

Abstract

We present an initial investigation into the performance of a multicollector inductively coupled plasmamass spectrometer equipped with a pre-mass filter (Neoma MC-ICP-MS/MS) for making plutonium (Pu) isotope ratio measurements on solutions containing low level (i.e., pg mL–1) Pu concentrations. This assessment was achieved by comparison of the 240Pu/239Pu, 241Pu/239Pu, and 242Pu/239Pu ratios attained over a one month period on the MC-ICP-MS/MS with the long-term (∼1 year) performance observed on the predecessor MC-ICP-MS (Neptune Plus) instrument each equipped with an equipped with an APEXΩ desolvating nebulizer for repeated measurements of certified reference materials from New Brunswick Program Office (NBL PO) CRM 136a and CRM 137. The MC-ICP-MS/MS performance of repeated measurement of CRM 136a (n = 20) resulted in mean values of 240Pu/239Pu = 0.1448 ± 0.0006, 241Pu/239Pu = 0.00371 ± 0.00006, and 242Pu/239Pu = 0.00682 ± 0.00006 (k = 2). The CRM 137 (n = 20), analyzed during the same analytical sessions, produced mean values for 240Pu/239Pu = 0.2414 ± 0.0006, 241Pu/239Pu = 0.00464 ± 0.00007, and 242Pu/239Pu = 0.0157 ± 0.0001 (k = 2). These results closely align with the certificate values for CRM 136a and CRM 137 and are within the k = 2 envelopes defined by the long-term performance of the traditional MC-ICP-MS approach (Neptune Plus). Examination of the performance of the various Pu isotope ratios as a function of total Pu content revealed accurate results (<3% relative difference, or RD) above ∼50 fg total Pu. The results presented here demonstrate the capability of the MC-ICP-MS/MS making accurate and precise low level Pu isotopic measurements. While the intent of this work was not to investigate the functionality of the collision cell, the pre-mass filter was employed. Future studies are warranted to investigate the entire capability of the MC-ICP-MS/MS collision cell and pre-cell mass filter optimization for performing low level Pu isotope measurements, even in mixed matrix samples.